Sevoflurane low-flow anaesthesia: best strategy to reduce Compound A concentration.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 12190805.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To define the best strategy to reduce Compound A production in Sevoflurane low-flow anaesthesia by experiments in vitro and in vivo of different absorbers and different anaesthesia machines. In vitro Compound A has been measured at 45 degrees C in vitro following Sevoflurane interactions with potassium hydroxide, sodium hydroxide, soda lime, Dragersorb 800 Plus and Amsorb, a new absorber that does not contain sodium or potassium hydroxide. In vivo Compound A concentration in the anaesthesia circuit (inspiratory branch) has been measured using an indirect sampling method through absorber vials (SKC) with active coal granules, during low flows (500 ml/min) general anaesthesia using soda lime, Dragersorb 800 Plus or Amsorb as absorber. Compound A was also measured during low flows (500 ml/min) general anaesthesia using as carbon dioxide absorber soda lime with different anaesthesia machines. In vitro at 45 degrees C Compound A concentration with soda lime and Dragersorb 800 Plus was about 10 times higher than with Amsorb. In vivo the Compound A concentrations in the inspiratory branch of the circuit were lower in the group with Amsorb. The Compound A production is minimal with Amsorb as carbon dioxide absorber.
Medical subject headings
- Anesthesia, Closed-Circuit
- Anesthetics, Inhalation
- Ethers
- Hydrocarbons, Fluorinated
- Methyl Ethers