Reproducibility of gait analysis variables: one-step versus three-step method of data acquisition.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 12194509.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of the study was to evaluate the preciseness and repeatability of EMED-SF platform data collection using two different protocols. Gait variables of five healthy women and five men, with an average age of 27.3 +/- 3.2 years, weighing 67.5 +/- 13.3 kg, were evaluated. With a one-step and a three-step approach of data collection, peak pressure, pressure-time integral, and contact time were measured. Peak pressures were not significantly different between both methods. Significant differences were found between both methods in total contact time and pressure-time integral. Both methods were comparable in peak pressures (error between methods = 7.0), while the one-step protocol was more repeatable (intraclass correlation coefficient = 0.59) than the three-step protocol (intraclass correlation coefficient = 0.36). The error between methods for total contact time and pressure-time integral were 143.3 and 50.1, respectively, suggesting that the two protocols lead to different results. The one-step protocol (intraclass correlation coefficient = 0.40) had a higher repeatability than the three-step protocol (intraclass correlation coefficient = 0.31). The one-step protocol has some advantages over the three-step protocol as far as repeatability, simplicity, convenience, and time conservation are concerned. For measuring total contact time and pressure-time integrals, both methods have comparable repeatability, although the protocols lead to different outcomes.
Medical subject headings
- Data Collection
- Diabetic Foot
- Foot
- Gait
Anatomy
- foot