Breast augmentation and reconstructive surgery: MR imaging of implant rupture and malignancy.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 12195470.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to assess the diagnostic accuracy of MRI in detecting prosthesis integrity and malignancy after breast augmentation and reconstruction. Forty-one implants in 25 patients were analyzed by MRI before surgical removal. Imaging results were compared with ex vivo findings. Magnetic resonance imaging of the breast was performed on a 1.5-T system using a dedicated surface breast coil. Axial and sagittal T2-weighted fast spin-echo as well as dynamic contrast-enhanced T1-weighted gradient-recalled-echo sequences were acquired. The linguine sign indicating collapse of the silicone shell or siliconomas indicating free silicone proved implant rupture, whereas early focal contrast enhancement of a lesion was suspicious for malignancy. The sensitivity for detection of implant rupture was 86.7% with a specificity of 88.5%. The positive and negative predictive values were 81.3 and 92.0%, respectively. The linguine sign as a predictor of intracapsular implant rupture had a sensitivity of 80% with a specificity of 96.2%. Magnetic resonance imaging revealed two lesions with suspicious contrast enhancement (one carcinoma, one extra-abdominal fibromatosis). Magnetic resonance imaging is a reliable and reproducible technique for diagnosing both implant rupture and malignant lesions in women after breast augmentation and reconstruction.
Medical subject headings
- Breast Implants
- Breast Neoplasms
- Magnetic Resonance Imaging