Digital subtraction in gadolinium-enhanced MR imaging of the brain: a method to reduce contrast dosage.

Chan, J H M; Tsui, E Y K; Chan, C Y; Lai, K F; Chau, L F; Fong, D; Mok, C K; Cheung, Y K et al. · Eur Radiol · 2002

prospective_cohort · Level II

Where this comes from

Abstract

The aim of the study was to investigate the feasibility of using digital subtraction in contrast-enhanced MR imaging of the brain to reduce the MR contrast dosage without jeopardizing patient care. Fifty-two patients with intracranial lesions, either intra-axial or extra-axial, detected by computerized tomography were selected for contrast-enhanced MR imaging with half-dose and full-dose of gadopentetate dimeglumine. The half-dose unsubtracted, full-dose unsubtracted, and half-dose subtracted MR images were visually assessed by counting the number of enhancing brain lesions in the images and quantitatively analyzed by computing their lesion contrast-to-background ratios (CBR). The visual conspicuity of the half-dose subtracted MR images was comparable to that of the full-dose unsubtracted MR images ( p>0.05), whereas the CBR of the half-dose subtracted images was approximately two to three times higher than that of the full-dose unsubtracted images. The half-dose subtracted T1-weighted spin-echo images might be able to replace the conventional standard-dose T1-weighted spin-echo images in MR imaging of the brain.