Augmented reality guidance for needle biopsies: an initial randomized, controlled trial in phantoms.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 12270235.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We report the results of a randomized, controlled trial to compare the accuracy of standard ultrasound-guided needle biopsy to biopsies performed using a 3D Augmented Reality (AR) guidance system. A board-certified radiologist conducted 50 core biopsies of breast phantoms, with biopsies randomly assigned to one of the methods in blocks of five biopsies each. The raw ultrasound data from each biopsy was recorded. Another board-certified radiologist, blinded to the actual biopsy guidance mechanism, evaluated the ultrasound recordings and determined the distance of the biopsy from the ideal position. A repeated measures analysis of variance indicated that the head-mounted display method led to a statistically significantly smaller mean deviation from the desired target than did the standard display method (2.48 mm for control versus 1.62 mm for augmented reality, p<0.02). This result suggests that AR systems can offer improved accuracy over traditional biopsy guidance methods.