Simplified calibration of single-plunge bipolar electrode array for field measurement during defibrillation.
other
Where this comes from
- Record sourced from PubMed, PMID 12374347.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In an earlier study, the authors presented a calibration technique for a triaxial bipolar electrode array (EA) that used 72 data points collected during a global sweep of the electric field vector relative to the EA axes. Although necessary for the initial characterization of the EAs, this data requirement has to be significantly reduced for the technique to become a practical tool. Therefore, in the present study, an analysis is performed to determine the relation between the number of data points used in the calibration and the mean root-mean-square error. The analysis shows that 18 data points can produce results nearly identical to those obtained with the 72-point calibration, thus reducing the required amount of data fourfold.
Medical subject headings
- Electric Countershock
- Electrocardiography
- Electrodes
- Electromagnetic Fields
- Models, Cardiovascular