A retrospective clinical and radiographic review of 173 hydroxyapatite-coated screw cups with 5- to 10-year follow-up, showing low revision rates for fixation failure.

Mann, C J; McNally, S; Taylor, E; Shepperd, J A N · J Arthroplasty · 2002

retrospective_cohort · Level III

Where this comes from

Abstract

We reviewed the midterm results of 173 hydroxyapatite-coated screw cups. The average follow-up was 6.5 years (range, 5-9 years). The follow-up rate was 93%. Patients were assessed using the Merle D'Aubigne-Postel clinical scoring scale and by radiographic review. Two patients had revision surgery for recurrent dislocation (1.2%), 3 patients were revised for aseptic loosening (1.7%), 1 patient underwent revision surgery because of deep prosthetic infection (0.6%), and 2 patients were revised for polyethylene wear without loosening (1.2%), which gave a total revision rate of 4.7%. The average postoperative Merle D'Aubigne-Postel scores were 5.7 for pain, 5.5 for range of motion, and 5.4 for function.

Medical subject headings

Anatomy