Direct plain radiographic methods versus EBRA-Digital for measuring implant migration after total hip arthroplasty.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 12375253.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We aimed to determine whether the precision and sensitivity of migration measurements after total hip arthroplasty (THA) using direct plain radiographic techniques could be made comparable to those of digital methods (EBRA-Digital; University of Innsbruck, Innsbruck, Austria) by careful control of radiographic technique and use of modern measuring tools. Precision was examined by analysis of consecutive radiographs taken after repositioning in 20 patients after hybrid THA. The precision (95% confidence interval) of measurements for cup migration using direct methods was +/-1.11 to 3.07 mm (x-axis) and +/-1.28 to 1.92 mm (y-axis). The precision of EBRA for cup measurements was +/-1.00 mm (x-axis) and +/-0.82 mm (y-axis). The precision of stem y-axis migration measurements was +/-1.12 to 6.91 mm using direct methods and +/-0.80 mm using EBRA. Migration of the stem (1.53 mm subsidence; P<.01) and the cup (0.53 mm cranial migration, P<.05) was detected using EBRA in 10 patients followed for 6 months after hybrid THA, but significant migration was not detectable using the most precise of the direct methods. Careful measures to standardize plain radiographs improve precision of direct radiographic measurements; however, their long-term sensitivity remains inferior to methods that employ quality control and measurement algorithms to measure migration from digitized radiographs.
Medical subject headings
- Hip Joint
- Hip Prosthesis
- Prosthesis Failure
- Software
Anatomy
- hip