Screw migration from total knee prostheses requiring subsequent surgery.

Shah, Steven N; Schurman, David J; Goodman, Stuart B · J Arthroplasty · 2002

case_report · Level V

Where this comes from

Abstract

Complications in total knee arthroplasty directly related to hardware failure other than polyethylene wear are rare. We report 2 cases of symptomatic screw migration into the joint space from total knee prostheses. In the first case, a screw disengaged from a constrained condylar knee prosthesis. Arthroscopy using standard arthroscopy portals and a small arthrotomy were performed to remove the screw. In the second case, symptomatic screw disengagement and posterior migration from the tibial component of a posterior-stabilized prosthesis occurred. Revision with replacement of the polyethylene insert and locking screw was required.