Evaluation of 70/30 D,L-PLa for use as a resorbable interbody fusion cage.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 12401023.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Titanium lumbar interbody spinal fusion devices are reported to be 90% effective for single-level lumbar interbody fusion, although radiographic determination of fusion has been debated. Using blinded radiographic, biomechanic, histologic, and statistical measures, researchers in the present study evaluated a radiolucent 70/30 poly(L-lactide-co-D,L-lactide) (70/30 D,L-PLa) interbody fusion device packed with autograft or rhBMP-2 on a collagen sponge in 25 sheep at 3, 6, 12, 18, and 24 months. A trend of increased fusion stiffness, radiographic fusion, and histologic fusion was demonstrated from 3 months to 24 months. Device degradation was associated with a mild to moderate chronic inflammatory response at all postoperative sacrifice times.
Medical subject headings
- Intervertebral Disc Displacement
- Lumbar Vertebrae
- Osseointegration
- Polyesters
- Spinal Fusion
Anatomy
- lumbar spine