Physical and chemical limitations to preparation of beta radioactive stents by direct neutron activation.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 12423597.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Pure beta emitters are the sources of choice for intracoronary irradiations in restenosis prevention. In this work we reconsidered preparation of low activity 32P sources by ion-implantation of stable 31P into highly biocompatible pure titanium stents, followed by neutron activation. Gamma-spectrometrical analysis has shown that during activations with high thermal neutrons flux production of gamma-active long-lived contaminants is much beyond the dosimetrically acceptable limit, mainly due to the competing (n,p) reactions induced by the fast neutrons on isotopes of the bulk stent material, and to a lesser extent due to (n,gamma) reactions on chemical impurities. A potential applicability of this method for obtaining alternative beta radioactive stents is discussed.
Medical subject headings
- Brachytherapy
- Coronary Disease
- Neutrons
- Stents