Evaluation of 70/30 poly (L-lactide-co-D,L-lactide) for use as a resorbable interbody fusion cage.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 12449196.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Titanium lumbar interbody spinal fusion devices are reported to be 90% effective in cases requiring single-level lumbar interbody arthrodesis, although radiographic determination of fusion has been debated. Using blinded radiographic, biomechanical, histological, and statistical measures, researchers in the present study evaluated a radiolucent 70/30 poly(L-lactide-co-D,L-lactide) interbody fusion device packed with autograft or recombinant human bone morphogenetic protein-2 on a collagen sponge in 25 sheep at 3, 6, 12, 18, and 24 months. A trend of increased fusion stiffness, radiographic fusion, and histologically confirmed fusion was demonstrated at 3 months to 24 months postimplantation. Device degradation was associated with a mild to moderate chronic inflammatory response at all postoperative sacrifice times. Use of this material in interbody fusion may be a viable alternative to metals.
Medical subject headings
- Absorbable Implants
- Biocompatible Materials
- Polyesters
- Spinal Fusion
- Spinal Injuries
- Transforming Growth Factor beta
Anatomy
- lumbar spine