Simulated use of premixed 0.25% isoflurane in 50% nitrous oxide and 50% oxygen.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 12453923.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Isoflurane (0.25%) in premixed nitrous oxide and oxygen, 50/50, v/v (IN(2)O), has been suggested for pain relief in labour. Possible phase separation of the mixture was studied during simulated administration. A sinusoidal pump set at stroke volume of 2 litres and a rate of 20-22 bpm and cycling for 1 min in three was used to simulate breathing during the painful contractions of labour. The temperature inside a 10-litre capacity cylinder did not drecrease sufficiently to cause separation of the gas mixture. Temperature in the demand valve decreased to -15.5 degrees C and this caused a small amount of liquid formation within the valve. Accordingly, the inspired concentration during the first breath of mixture in a cycle could be transiently as high as 0.55%. The concentration observed at the patient connection after the first breath varied between 0.17 and 0.28%. The system delivered a clinically acceptable performance although further development to avoid liquid condensation is needed.
Medical subject headings
- Anesthetics, Combined
- Cold Temperature
- Isoflurane
- Nitrous Oxide
- Oxygen