Effect of metal and sampling rate on accuracy of Flock of Birds electromagnetic tracking system.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 12485650.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Electromagnetic tracking devices are used in many biomechanics applications. Previous studies have shown that metal located within the working field of direct current electromagnetic tracking devices produces significant errors. However, the effect of sampling rate on the errors produced in a metallic environment has never been studied. In this study, the accuracy of Ascension Technologies' Flock of Birds was evaluated at sampling rates of 20, 60, 100, and 140 Hz, in the presence of both aluminum and steel. Aluminum interference caused an increase in measurement error as the sampling rate increased. Conversely, steel interference caused a decrease in measurement error as the sampling rate increased. We concluded that the accuracy of the Flock of Birds tracking system can be optimized in the presence of metal by careful choice in sampling rate.
Medical subject headings
- Electromagnetic Phenomena
- Equipment Failure Analysis
- Metals
- Motion
- Sample Size
- Telemetry