A comparison of screening methods in school-aged children.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 12501102.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We sought to compare the findings of 3 different hearing screening methods in school-aged children. Prospective testing of second- and third-grade students in their schools was carried out. Three hundred children (599 ears) were screened by using 3 test modalities, pure-tone audiometry, distortion product otoacoustic emissions (DPOAE), and tympanometry. All of the tests were normal in 532 ears (89%), and all were abnormal in 12 ears (2%). Tympanometry yielded the most abnormalities (8.3%), and pure-tone testing demonstrated the fewest (3.3%), with a positive rate of 6.3% for DPOAE testing. False-positive rates were 1.2%, 4.2%, and 6.4% for pure tones, DPOAE, and tympanometry, respectively, when normal results on pure-tones or DPOAE were taken to represent true hearing. We continue to recommend pure-tone testing as an effective screening method, with follow-up by using otoacoustic emissions in those who fail the pure-tone test.
Medical subject headings
- Acoustic Impedance Tests
- Audiometry, Pure-Tone
- Hearing Disorders
- Mass Screening
- Otoacoustic Emissions, Spontaneous