Biomechanical influences of magnetic resonance imaging on the SOUNDTEC Direct System implant.

Dyer, Robert Kent; Dormer, Kennet J; Hough, J V D; Nakmali, Udon; Wickersham, Rick · Otolaryngol Head Neck Surg · 2002

biomechanical · Level V

Where this comes from

Abstract

The purpose of this study was to measure the forces experienced by the SOUNDTEC Direct System magnetic implant during 0.3-T MRI. Torsional and linear forces imposed on 8 implants were measured by using calibrated neurologic Von Frey hairs and were compared with finite-element analysis predictions and the forces required to separate the incudostapedial joints of unpreserved temporal bones. An implanted embalmed autopsy specimen was also examined before and after 1.5-T MRI. Peak linear force at the orifice of the MRI core measured 0.51g (+/-0.2 SD). Maximum torque occurred at the MRI core center and measured 11.4g-cm (+/-1.2 SD). The mean torque required to separate the incudostapedial joints of 12 unpreserved temporal bones was 33.8g-cm (+/-20.4 SD). The autopsy specimen sustained a 1.5-T MRI scan without disruption of the ossicular chain or explantation. Physical and mechanical testing of the SOUNDTEC implant indicates that the structural integrity of the ossicles will be maintained during 0.3-T MRI of the human head.

Medical subject headings