Lack of specificity of cyclosporine immunoassays. Results of a College of American Pathologists Study.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 12521361.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate the cross-reactivity of the 6 most abundant cyclosporine A (CsA) metabolites in commonly used assays for CsA. Whole blood samples containing either only 62 ng/mL CsA (A) or 62 ng/mL CsA and between 49 and 86 ng/mL of 1 of the 6 most abundant CsA metabolites (B) were lyophilized. One sample of A and 1 of B were mailed to each of the laboratories participating in the College of American Pathologists Proficiency Testing Program quarterly during a 3-year period (1999-2001). Method means and coefficients of variation were calculated for each mailing. The study showed significant cross-reactivity of metabolites in all the immunoassay systems studied. Overall degree of interference decreased from Abbott TDx polyclonal > Abbott TDx monoclonal > DiaSorin > Syva EMIT. High-performance liquid chromatography methods gave results close to those found using mass spectrometric techniques. Significant metabolite interference was found to occur with the immunoassay systems studied.
Medical subject headings
- Cyclosporine
- Immunoassay