Intermittent or continuous carbon dioxide insufflation for de-airing of the cardiothoracic wound cavity? An experimental study with a new gas-diffuser.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 12538172.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Insufflation of carbon dioxide into the chest wound is used in open-heart surgery to de-air the heart and great vessels. In a cardiothoracic wound model, we compared the degree of air displacement achieved by a new insufflation device, a gas-diffuser, with that of a thin open-ended tube during steady-state and with carbon dioxide flows of 2.5, 5, 7.5, and 10 L/min. We also studied air displacement at the start of and after discontinuation of carbon dioxide insufflation with the gas-diffuser and evaluated the influence of an open pleura. During steady state, the gas-diffuser produced efficient air displacement in the wound cavity model at carbon dioxide flows of > or = 5 L/min (< or = 0.65% remaining air), whereas the open-ended tube was inefficient (> or = 82% remaining air) at all studied carbon dioxide flows (P < 0.001). An open pleural cavity prolonged the time needed to obtain a high degree of air displacement in the wound cavity (P = 0.001). Carbon dioxide insufflation of the cardiothoracic wound cavity should be initiated at a carbon dioxide flow of 10 L/min at least 1 min before the incision of the heart and great vessels and should be continued at a carbon dioxide flow of at least 5 L/min until surgical closure.
Medical subject headings
- Carbon Dioxide
- Cardiac Surgical Procedures
- Insufflation