Calibration of three capnographs for use with helium and oxygen gas mixtures.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 12562412.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Capnography is considered essential in the management of mechanically-ventilated patients. Helium, as an adjunct to mechanical ventilation, is the subject of renewed interest and used increasingly. However, helium affects the performance of infrared capnometry. We constructed a simple device to generate variable mixtures of helium, oxygen and carbon dioxide within the normal physiological range, and tested the performance of two side-stream and one in-line capnographs. We found that addition of helium to the gas mixture caused all three capnographs to underestimate the concentration of carbon dioxide. The underestimation increased as the proportion of helium increased. The maximum underestimation (30%) occurred in a 79:21 helium/oxygen mixture.
Medical subject headings
- Capnography
- Carbon Dioxide
- Helium
- Oxygen Inhalation Therapy
- Respiratory Insufficiency