Resolution requirements for monitor viewing of digital flat-panel detector radiographs: a contrast detail analysis.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 12599009.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
With the introduction of digital flat-panel detector systems into clinical practice, the still unresolved question of resolution requirements for picture archiving communication system (PACS) workstation monitors has gained new momentum. This contrast detail analysis was thus performed to define the differences in observer performance in the detection of small low-contrast objects on clinical 1K and 2K monitor workstations. Images of the CDRAD 2.0 phantom were acquired at varying exposures on an indirect-type digital flat-panel detector. Three observers evaluated a total of 15 images each with respect to the threshold contrast for each detail size. The numbers of correctly identified objects were determined for all image subsets. No significant difference in the correct detection ratio was detected among the observers; however, the difference between the two types of workstations (1K vs 2K monitors) despite less than 3% was significant at a 95% confidence level. Slight but statistically significant differences exist in the detection of low-contrast nodular details visualized on 1K- and 2K-monitor workstations. Further work is needed to see if this result holds true also for comparison of clinical flat-panel detector images and may, for example, exert an influence on the diagnostic accuracy of chest X-ray readings.
Medical subject headings
- Computer Terminals
- Data Display
- Image Enhancement
- Radiology Information Systems