Noncontact radio-frequency ablation for obtaining deeper lesions.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 12665035.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Radio-frequency (RF) cardiac catheter ablation has been very successful for treating some cardiac arrhythmias, however, the success rate for ventricular tachycardias is still not satisfactory. Some existing methods for developing deeper lesions include active cooling of the electrode and modifying the electrode shape. We propose a method of noncontact ablation, to solve this problem. We apply 120 W of power through an 8-mm electrode for a 120-s duration, with distances from 0 to 3 mm between electrode and myocardium, to create lesions in myocardium. We apply flow rates of 1, 3, and 5 L/min to determine their effect. Results show that with an optimal distance from 0.5 to 1.5 mm between electrode and myocardium, we increase lesion depth from 7.5 mm for contact ablation to 9.5 mm for noncontact ablation. For different flow rates, the optimal distance various. The effect of flow rate is not obvious. Higher flow rate does not lead to a deeper lesion.
Medical subject headings
- Biosensing Techniques
- Catheter Ablation
- Heart Ventricles
- Radiofrequency Therapy