Experimental studies on multiple-model predictive control for automated regulation of hemodynamic variables.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 12669984.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A model-based control methodology was developed for automated regulation of mean arterial pressure and cardiac output in critical care subjects using inotropic and vasoactive drugs. The control algorithm used a multiple-model adaptive approach in a model predictive control framework to account for variability and explicitly handle drug rate constraints. The controller was experimentally evaluated on canines that were pharmacologically altered to exhibit symptoms of hypertension and depressed cardiac output. The controller performed better as compared to experiments on manual regulation of the hemodynamic variables. After the model bank was determined, mean arterial pressure was held within +/- 5 mm Hg 88.9% of the time with a standard deviation of 3.9 mm Hg. The cardiac output was held within +/- 1 l/min 96.1% of the time with a standard deviation of 0.5 l/min. The manual runs maintain mean arterial pressure only 82.3% of the time with a standard deviation of 5 mm Hg, and cardiac output 92.2% of the time with a standard deviation of 0.6 l/min.
Medical subject headings
- Blood Pressure
- Cardiac Output
- Cardiovascular Diseases
- Dopamine
- Infusion Pumps
- Models, Cardiovascular
- Phenylephrine