Estimation of cardiac conduction velocities using small data sets.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 12680723.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Cardiac surface conduction velocities could provide valuable information about the speed and angle of a propagating electrical wave front. It would be advantageous to develop catheter-based velocity mapping devices to improve the visualization of cardiac arrhythmias. However, catheter tip size is limited to < or = 2.5 mm in diameter, restricting the number and size of electrodes that can be placed on a catheter tip. We address the feasibility of estimating conduction speed and angle from small data sets suitable for recording from a catheter device in a standard clinical environment. We estimated cardiac conduction velocities from data subsets of 4-7 closely spaced electrograms, and then compared these estimates to velocities estimated from a larger reference grid. We studied 137 ventricular beats and 17,756 velocity vectors from six swine hearts. Average differences in angle between the two estimates were 0.4 degrees +/- 16 degrees while average differences in speed were 5% +/- 33%. These angle and speed differences provide an initial quantitative assessment of velocity accuracy for the purposes of catheter-based vector mapping.
Medical subject headings
- Body Surface Potential Mapping
- Heart Conduction System
- Models, Cardiovascular
- Neural Conduction