Biomechanical evaluation of a bipedicular spinal fixation device: three different strength tests.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 12687438.
- Also identified by PMC identifier 3468004.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Disadvantages of thoracic posterior implants and developments in rod contouring in situ led to the design of a new spinal implant: the bipedicular spinal fixation device (BSF). The BSF is composed of two bifid hooks linked by a compression transverse connector and inserted into the costo-vertebral and costo-transverse joints. The aim of this biomechanical study was to determine the loading tolerance of the BSF. Three strength tests-a pull-out test, a lateral load-to-failure test and a uniaxial transversal compression test to failure-were performed using six human thoracic spines on an Instron testing device. Specimen evaluation consisted of: bone mineral density (BMD) measurement with the dual-energy X-ray absorptiometry (DEXA) technique, cortical thickness measurements and a morphometric study. The mean values for load-to-failure in the posterior and lateral tests were 324 N and 400 N respectively. The mean value of the uniaxial compression was 988 N. The mean BMD estimated by DEXA was 0.557 g/cm(2). The BSF loading tolerance was compatible with the in situ rod contouring technique requirements when we considered posterior and lateral pull-out tests. The transversal compression test determined the appropriate and efficient BSF tightening force.
Medical subject headings
- Internal Fixators
- Materials Testing
- Spinal Fusion
- Thoracic Vertebrae
Anatomy
- thoracic spine