Validation of a tonometric noninvasive arterial blood pressure monitor in the intensive care setting.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 12694001.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Intra-arterial measurement is considered the gold standard for continuous, beat-to-beat arterial blood pressure monitoring. However, arterial cannulation can be difficult and may cause complications such as thrombosis and ischaemia. Recently, a tonometric system, the Colin CBM-7000 has been developed for noninvasive beat-to-beat measurement of arterial blood pressure from the radial artery. We assessed the level of agreement between the CBM-7000 and invasive radial artery measurements in 15 patients on a neuro-intensive care unit. Agreement of systolic, diastolic and mean arterial pressure values was limited, with approximately 34% of mean arterial pressures differing by over 10 mmHg. In many cases, this was due to a downward drift of the noninvasive measurements over time. Furthermore, there was a tendency to underestimate low pressures and overestimate high pressures. In our opinion, the Colin CBM-7000 cannot be recommended for continuous blood pressure monitoring in the intensive care setting.
Medical subject headings
- Blood Pressure Monitors
- Critical Care