The use of a constrained acetabular component for recurrent dislocation.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 12728414.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The poor results of surgical treatment of chronic instability after total hip arthroplasty (THA) led to the development of a constrained acetabular component. In this study, 87 constrained THAs implanted for recurrent instability were reviewed retrospectively. Eighty-five hips were available for follow-up evaluation, with an average follow-up period of 58 months. These 85 hips were evaluated at a minimum of 3 years. Two recurrent dislocations were seen, caused by dissociation of the liner from the shell. Four acetabular components and 1 femoral component were revised. Overall, a 2.4% dislocation rate and an 8.2% revision rate were seen. The recurrent dislocation rate of 2.4% represents a significant improvement over other methods reported. Repeat dislocation was only seen in dissociation of cemented liners into well-fixed shells. We do not recommend this mode of fixation.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Dislocation
- Hip Prosthesis
- Prosthesis Failure