Primary hybrid total hip arthroplasty with a roughened femoral stem: integrity of the stem-cement interface.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 12728421.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
One hundred and two consecutive cemented femoral stems were evaluated in 92 patients at an average 9-year follow-up and a minimum 5-year follow-up (range, 5-14 years). The stem used was cobalt chromium with a collar, normalization steps, and a roughened surface (Ra 40); the stem was inserted using contemporary cementing techniques. This series demonstrated a femoral component aseptic loosening rate of 2.0% and a femoral component survivorship of 97.2 +/- 2.0% at 10 years. One of 2 failed stems was revised at 95 months for failure at the cement-bone interface. The second failed stem showed failure at the cement-bone interface with incomplete debonding radiographically at 65 months. The remaining femoral components did not demonstrate any evidence of debonding at the stem-cement interface. These results compare favorably with other series of cemented femoral stems, as well as with those with a polished surface.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Cementation
- Hip Prosthesis
- Prosthesis Failure
Anatomy
- femur
- hip