Unexpected durability of Smeloff-Cutter aortic ball valve prosthesis.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 12735595.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We report a case in which replacement of a Smeloff-Cutter aortic ball prosthesis was required 28 years after initial implantation. A 57-year-old woman underwent aortic valve replacement with a 21-mm Smeloff-Cutter ball prosthesis and open mitral commissurotomy for aortic stenosis, aortic regurgitation, and mitral stenosis in 1973. Severe aortic regurgitation occurred in April 2001, and aortic valve reoperation combined with mitral valve replacement was successfully performed. The patient's aortic ball valve was nearly intact with perivalvular leakage probably causing the aortic regurgitation. Our experience documents longer durability for the Smeloff-Cutter prosthesis than has been reported to date.
Medical subject headings
- Aortic Valve
- Heart Valve Prosthesis
- Heart Valve Prosthesis Implantation
- Prosthesis Design