Intraoperative high-dose-rate 192Ir radical implant in early breast cancer: a quality assurance and dosimetry study.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 12788174.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate the variability in catheter length, geometry, and dosimetric parameters of radical intraoperative high-dose-rate breasts implant during 7-11 days. Simulator X-rays, CT scans, and dosimetric studies were repeated on alternate days in 14 consecutive patients treated with radical intraoperative two- or three-plane nylon catheter high-dose-rate implant (34 Gy in 10 fractions within 5 days). Significant variation was found in catheter length, but no major change was noted in implant geometry, homogeneity, or inhomogeneity indexes. A variation in length of >5 mm in one or more catheters was seen in all patients and >10 mm in 11 patients at any time during the implant. Of the 171 catheters in 14 patients, 100 (58%) and 38 (22%) showed a variation of >5 mm or >10 mm, respectively. The variation of >10 mm was reduced from 32% of catheters in the first 5 patients to 17% in the subsequent 9 patients (p = 0.028). Rigid catheter fixation might reduce length variation but may cause skin necrosis if the expanding cavity indents the skin for a long period against the fixation device. A dose homogeneity index of 0.90 (range 0.85-0.92) and dose nonuniformity ratio of 0.20 (range 0.12-0.25) were satisfactory. The catheter fixation and exit catheter length should be measured daily and if the implant is in situ for more than a few days, orthogonal X-rays and, if indicated, dosimetry should be repeated at least once.
Medical subject headings
- Brachytherapy
- Breast Neoplasms
- Iridium Radioisotopes