The rigid nasendoscope as a tool for difficult tracheal intubation: a manikin study.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 12790816.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We examined the use of the 30 degrees rigid nasendoscope in aiding difficult tracheal intubations. A Cormack and Lehane grade 4 difficult intubation (no view of glottis or epiglottis) was set up on a manikin. After 10 s of tuition, 40 anaesthetists attempted to pass a standard gum elastic bougie between the cords, with and without the nasendoscope, in randomised order. A bougie curved to an 'optimal curve' was also tested. Using the standard bougie 13/40 (33%) passed the bougie between the cords without the nasendoscope, compared with 31/40 (78%) when using the nasendoscope (p < 0.001). The 'optimal curve' bougie resulted in 29/40 (73%) and 39/40 (98%) success rates without and with the nasendoscope, respectively (p = 0.004). The nasendoscope is a simple and easy to use tool in grade 4 intubation, and results are improved further by the use of an 'optimal curve' bougie.
Medical subject headings
- Endoscopes
- Intubation, Intratracheal