Myotonometer intra- and interrater reliabilities.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 12808553.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To assess the intra- and interrater reliabilities of the Myotonometer, a hand-held, computerized, electronic device that quantifies muscle stiffness (tone/compliance). Reliability study. Research laboratory. Thirty-five healthy, nondisabled adults (age range, 22-42 y). Not applicable. Two raters used the Myotonometer to evaluate subjects' lateral gastrocnemius and biceps brachii muscles. Muscles were measured in a relaxed state and during a voluntary isometric contraction. Coefficients were calculated for each muscle and each condition (relaxed, contracted). Results were analyzed by using Design II intraclass correlation coefficients. Reliability coefficients were highest when the instrument exerted moderate to strong forces against the muscle (range, 0.50-2.00 kg; intrarater reliability R range, .84 - .99; interrater reliability R range, .75 - .96). Myotonometer measurements had high to very high intra- and interrater reliabilities for measurements of the lateral gastrocnemius and biceps brachii muscles.
Medical subject headings
- Manometry
- Muscle Tonus
- Muscle, Skeletal