Body contour 180 degrees pinhole SPET with or without tilted detector: a phantom study.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 12811530.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study investigated the feasibility of ordered subsets expectation maximisation (OS-EM) reconstruction of 180 degrees pinhole single-photon emission tomography (SPET) acquired in body contour mode (variable distance between the detector and the axis of rotation for each projection) with or without a tilted detector head. Four non-circular orbits were designed bearing in mind the rotation radius and tilt angle values of previous pinhole SPET acquisitions in patients with circular orbits. The reconstructions were performed using a dedicated OS-EM algorithm. Reconstructed images of line and uniformity phantoms showed that the spatial and uniformity characteristics of the radioactive objects were preserved. In comparison with the circular orbits, the non-circular orbits allowed only a moderate gain (maximum 10%) in resolution. However, body contour pinhole SPET would significantly facilitate the camera set-up and in this way should decrease the camera set-up time, which is an important parameter in patient studies.
Medical subject headings
- Gamma Cameras
- Image Enhancement
- Imaging, Three-Dimensional
- Phantoms, Imaging
- Tomography, Emission-Computed, Single-Photon