Clinical field-testing of ATLAS prosthetic system for trans-femoral amputees.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 12812328.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The ATLAS prosthetic system was tested on 47 trans-femoral amputees in two tropical areas. Although the craftsmanship was not acceptable, 46% (16/35) were malaligned many with the foot dorsiflexed, the user compliance was good (91%, 32/35), and the satisfaction rate high (71%, 25/35). However, after 7-10 months failures occurred in 83% (29/35) mostly because of problems with the ATLAS knee mechanism. A modification was tested and 80% (12/15) had again failed at the knee after 6 (2-12) months. The ATLAS system in its current design should be abandoned for use for trans-femoral amputees.