Carotid angioplasty in a pulsatile flow model: factors affecting embolic potential.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 12819644.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Carotid endoluminal intervention is an alternative to surgery but carries a risk of embolic stroke even with distal protection devices. We investigated the clinical features and degree of stenosis related to number and size of emboli during carotid angioplasty. An experimental ex vivo study. An ex vivo pulsatile flow model was used in which temperature, velocity, flow, pressure and viscosity characteristics were designed to simulate the carotid circulation. Carotid endarterectomy specimens excised as intact cylinders (n = 28) were subjected to a standardised angioplasty procedure using radiological guidance. Emboli collected in filters placed distally were counted and sized using microscopy. Median number of emboli during angioplasty was 133 (range 15-1331). Median size of the largest embolus was 700 microns (range 75-2400). Severity of stenosis correlated with increased maximum size (r = 0.55, p = 0.012). Statin therapy >4 weeks pre-operatively was associated with reduced emboli number and size (54 (range 15-748) vs 247 (range 37-1331) [p = 0.023] and 400 microm (range 75-2400) vs 1300 microm (range 600-2200) [p = 0.022]). In this model a wide range of emboli number and size were produced. Number and size of embolic particles were highest in patients with high-grade stenoses not receiving statin therapy.
Medical subject headings
- Carotid Stenosis
- Embolism
- Endarterectomy, Carotid
- Pulsatile Flow