Constrained acetabular cups: a cadaveric biomechanical evaluation.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 12820090.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Reports indicate that constrained acetabular cups may reduce range of motion and catastrophically fail, although the biomechanics of dislocation have not been reported. We measured the available motion in 6 constrained cups (from 2 manufacturers) when anatomically placed in cadaver pelves. We measured the torque and rotation necessary to dislocate the hip. Range of motion was measured using a custom jig and revealed a functional range of motion with all cup positions. Extension was limited to 0 degrees with anteversion of 11 degrees to 29 degrees. Thus, anteversion of <10 degrees is recommended. Torque and rotation to produce dislocation was 7 ft-pounds and 11 ft-pounds and 82 degrees and 72 degrees, for each manufacturer. Torques were reduced with subsequent dislocations. We recommend consideration of polyethylene replacement if dislocation occurs. Dislocation occurred at the ball-polyethylene interface without catastrophic failure.
Medical subject headings
- Hip Dislocation
- Hip Prosthesis
Anatomy
- hip
- pelvis