Evaluation of a new CPC-to-gauge bonding technique with the use of in vitro fluid flow.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 12861601.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Strain gauging enables the measurement of bone deformation during physical activity, leading to a better understanding of the physiological effects of loading on bone growth and remodeling. Development of a technology that will withstand long-term in vivo exposure and bond securely to bone is imperative for accurate, consistent measurement collection. Polysulfone is currently used to attach calcium-phosphate ceramic (CPC) particles, which promote bone-to-gauge bonding, to polyimide-backed strain gauges. This study evaluated the use of an implant-grade epoxy as an alternative CPC-polyimide adhesive. Polyimide-epoxy-CPC interfaces were loaded to failure and shear strengths calculated. In vitro studies providing a constant flow of medium over test specimens were designed, and long-term in vitro fluid exposure studies of the epoxy's shear strength were conducted. Average shear strength of polysulfone-polyimide interfaces were reported to be 7 MPa. The average shear strength of the epoxy-polyimide interface before long-term in vitro exposure was 17 MPa, which is stronger than the shear strength of the bone-CPC interface. The strength of the epoxy-polyimide interface decreased to 6.8 MPa after 24 weeks in vitro and 3 MPa after 24 weeks in vivo.
Medical subject headings
- Biocompatible Materials
- Bone and Bones
- Calcium Phosphates
- Ceramics
- Epoxy Compounds