Inadvertent positive end-expiratory pressure caused by a malfunctioning ventilator relief valve.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 12873943.
- Also identified by DOI 10.1213/01.ANE.0000072546.38277.10.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
During anesthesia, a nurse draped several gas hoses over the ventilator relief valve on the back of the anesthesia machine, causing a malfunction. Unintended positive end-expiratory pressure was administered to the patient. Causes of this mishap, anesthesia machine design, and nonanesthesiologist familiarity with anesthesia machine components and their function, are discussed.