A dual optic accommodating foldable intraocular lens.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 12928270.
- Also identified by PMC identifier 1771831.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To design an accommodating intraocular lens with extended accommodative range that can be adapted to current standard phacoemulsification and endocapsular implantation technique. Ray tracing analysis and lens design; cadaver eye implantation. Ray tracing analysis indicated that axial movement of an exaggerated converging anterior optic linked by spring loaded haptics to a compensatory static diverging posterior optic produced greater change in conjugation power of the eye compared to axial movement of a single optic lens. A dual optic one piece foldable silicone lens was constructed and implanted via a 4 mm corneal incision into the capsular bag of two cadaver eyes. A dual optic intraocular lens design can increase the optical effect of a given displacement and suggests improvements for accommodating intraocular lenses.
Medical subject headings
- Lenses, Intraocular
- Optics and Photonics
- Prosthesis Design