A modified ultrasensitive assay to detect quantified HIV-1 RNA of fewer than 50 copies per milliliter.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 12931558.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We compared the sensitivity and specificity of versions 1.0 and 1.5 and a modified version 1.5 of the AMPLICOR HIV-1 MONITOR ultrasensitive RNA assay (Roche, Indianapolis, IN) by using a virus stock dilution series and plasma samples from HIV-1-infected and uninfected subjects. The modified assay was linear and consistently positive down to 12 copies per milliliter vs 25 copies per milliliter for the other 2 assays. Versions 1.0, 1.5, and modified 1.5, respectively, detected 9 (23%) of 39, 11 (28%) of 40, and 43 (61%) of 71 replicates of a 4-copy-number and 11 (28%) of 40, 17 (46%) of 37, and 88 (90%) of 98 replicates of a 10-copy-number standard. Of 44 patient samples with undetectable levels using version 1.0, 32 (73%) had detectable levels on the modified assay, and 5 (25%) of 20 had detectable levels on version 1.5. None of the assays detected HIV-1 RNA in HIV-1 antibody-negative samples. The modified version 1.5 of the RNA assay is more sensitive for detecting HIV-1 RNA in significantly more patients than are versions 1.0 and 1.5.
Medical subject headings
- Gene Dosage
- HIV Infections
- HIV-1
- RNA, Viral
- Viremia