A comparative evaluation of transcutaneous and end-tidal measurements of CO2 in thoracic anesthesia.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 12933401.
- Also identified by DOI 10.1213/01.ANE.0000074793.12070.1E.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We performed this study to assess the accuracy of transcutaneous CO(2) (PTCCO(2)) monitoring compared with end-tidal CO(2) (PETCO(2)) in thoracic anesthesia. Twenty-six patients undergoing pneumonectomy with thoracotomy for which a long period of one-lung ventilation (OLV) was required were studied. The lungs were mechanically ventilated in the lateral decubitus position. PTCCO(2), PETCO(2), and arterial CO(2) (PaCO(2)) were simultaneously measured during two-lung ventilation (TLV) and during OLV at intervals of 15 min. All patients completed the study protocol. Bland-Altman analysis revealed a bias of -0.4 mm Hg with a precision of +/-2.5 mm Hg during OLV and 1.4 mm Hg with +/-4.3 mm Hg during TLV when PTCCO(2) and PaCO(2) were compared and revealed a bias of -5.8 mm Hg with a precision of +/-4.1 mm Hg during OLV and -7.1 mm Hg with +/-4.6 mm Hg during TLV when PETCO(2) and PaCO(2) were compared. We conclude that PTCCO(2) monitoring is accurate for evaluating CO(2) levels during thoracic anesthesia.
Medical subject headings
- Anesthesia
- Carbon Dioxide