Silicone gel infiltration of a peripheral nerve and constrictive neuropathy following rupture of a breast prosthesis.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 1313982.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Following rupture of a subpectoral breast prosthesis, massive amounts of silicone gel migrated into the arm of a patient. The patient developed painful paresthesias and decreased sensation in the cutaneous distribution of the superficial radial nerve. Nerve conduction studies showed both an increase in distal latency and decreased amplitude in this nerve compared with the normal opposite side. Subsequent neurolysis confirmed dense fibrosis surrounding the nerve. Silicone droplets also were observed within the thickened epineurium of the median nerve, but no electrophysiologic evidence of neuropathy occurred. Multiple debridements of the subcutaneous tissue of the arm were necessary. In one of these specimens, histologic sections demonstrated silicone gel infiltration of a subcutaneous nerve. This is the first reported case of silicone gel infiltration of a nerve and constrictive neuropathy associated with a prosthesis rupture.
Medical subject headings
- Nerve Compression Syndromes
- Prostheses and Implants
- Radial Nerve
- Silicones
Anatomy
- radius