Polyethylene debris-induced osteolysis and loosening in uncemented total hip arthroplasty. A cause of late failure.
review · Level V
Where this comes from
- Record sourced from PubMed, PMID 1402944.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Uncemented total hip arthroplasty has proven to be an acceptable alternative to cemented total hip arthroplasty with good short-term results. With the elimination of the use of polymethyl methacrylate for component fixation, failure at the bone-cement interface, with resultant osteolysis and progressive loosening, was thought to be preventable. Unfortunately the ultra-high-molecular-weight polyethylene acetabular insert can wear and produce particulate debris. This debris can stimulate an osteolytic reaction and lead to late aseptic loosening in a cementless total hip arthroplasty.
Medical subject headings
- Hip Prosthesis
- Osteolysis
- Polyethylenes