The electrophysiologic and histologic effects of the Argon Beam Coagulator on peripheral nerves.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 1405988.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The electrophysiologic and histologic effects of the Bard Argon Beam Coagulator (ABC) were investigated in the New Zealand White rabbit. Thirty-four rabbits were divided into three groups. Controls underwent simple femoral exploration and closure. The remaining rabbits' femoral nerves were spot coagulated with either the ABC or standard electrosurgical unit (ESU). Stimulus thresholds were recorded before treatment and again prior to sacrifice at 0, 30, 60, or 120 days. Thresholds were significantly elevated for the ABC and ESU compared to controls (P = .0077 and .0351, respectively). Changes in threshold were greater for the ABC than for the ESU, but were not significant. All ABC- and ESU-treated nerves had significant histologic injury when compared to controls (P < .0002). Although the ABC may be clinically safe, significant injury to rabbit femoral nerves occurs when they are exposed to energy emitted by this instrument.
Medical subject headings
- Argon
- Electrocoagulation
- Femoral Nerve