Amniotic fluid volume assessment: comparison of ultrasonographic estimates versus direct measurements with a dye-dilution technique in human pregnancy.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 1415438.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of our study was to compare the accuracy of clinical ultrasonographic techniques of amniotic fluid volume assessment with a dye-dilution technique. We compared amniotic fluid volume as measured by ultrasonographic techniques with a dye-dilution method in 50 women undergoing amniocentesis during the third trimester. Thirteen separate ultrasonographic techniques, including the amniotic fluid index, were evaluated with regression analysis. Amniotic fluid volumes as determined by dye-dilution ranged between 129 and 4444 ml. The amniotic fluid index overestimated the actual volume by as much as 88.7% at lower volumes and underestimated the actual volume by as much as 53.9% at higher volumes. Differences in measurement error between the other ultrasonographic methods and the amniotic fluid index did not appear to be sufficient to warrant changes in current clinical practice. A major source of error in ultrasonographic amniotic fluid volume assessment is that one-dimensional measurements are used to estimate the volume of a complex, three-dimensional object.
Medical subject headings
- Amniotic Fluid
- Indicator Dilution Techniques
- Ultrasonography, Prenatal