Fast spin-echo (FSE) magnetic resonance (MR) imaging was compared with conventional, peripherally gated T2-weighted spin-echo (SE) imaging in the detection of high- and low-signal-intensity lesions in the central nervous system. Lesion detectability was determined with percentage of contrast measurements and contrast-to-noise ratios with two different measurements for noise. All three measures of lesion detectability were similar. FSE and SE sequences were quantitatively equivalent in the detection of high-signal-intensity lesions. The SE sequence, however, was superior to the FSE sequence in the detection of small, low-signal-intensity lesions in the central nervous system caused by magnetic susceptibility effects.
Medical subject headings
- Brain Diseases
- Magnetic Resonance Imaging