Simultaneous measurement of preductal and postductal oxygen saturation by pulse oximetry in hyaline membrane disease.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 1444550.
- Also identified by PMC identifier 1590451.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Preductal and postductal oxygen saturation were compared in 20 ventilated preterm infants with hyaline membrane disease to establish the frequency of right to left shunting and to assess the accuracy of postductus arteriosus blood gas sampling. One hundred and thirty eight comparisons were made and the frequency of right to left shunting was 17% (95% confidence interval 12 to 25%). Shunting episodes with possible preductal hyperoxia occurred far less commonly on a maximum of 5% of occasions. The findings in infants under 1000 g and of 24-28 weeks' gestation were not significantly different from larger or more mature infants. Shunting occurred significantly more frequently in very ill infants who subsequently died as a result of respiratory disease.