Is it advantageous to strengthen the cement-metal interface and use a collar for cemented femoral components of total hip replacements?

Harris, W H · Clin Orthop Relat Res · 1992

review · Level V

Where this comes from

Abstract

The mechanism of initiation of loosening of cemented femoral components is now known. It is debonding at the cement-metal interface. Current data strongly support the concept that a collar and improvement of the cement-metal interface are valuable. Precoating and having a roughened surface proximally and distally on the stem contribute to extended longevity of the cement-metal interface. Using contemporary instrumentation, collar-calcar contact can be achieved regularly and, once achieved, is well maintained for years.

Medical subject headings

Anatomy