Rapid fetal lung maturity testing: commercial versus NBD-phosphatidylcholine assay.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 1448250.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We compared the TDx Fetal Lung Maturity test and the fluorescence polarization method using 1-palmitoyl-2(6-[(7-nitro-2,1,3-benzoxadiazol-4- yl)amino]caproyl)phosphatidylcholine (NBD-phosphatidylcholine). Using 76 paired human amniotic fluid samples, the fluorescence polarization values of the two methods were found to have a strong nonlinear correlation (r2 = 0.946). Both assays can be completed in less than 1 hour, have excellent precision (between-day variation less than 2%), and indicate the amount of surfactant phospholipid relative to albumin. The FLM assay is calibrated with surfactant/albumin standards; therefore, the reported results (in mg/g) correlate inversely with polarization of NBD-phosphatidylcholine. Strong correlations were seen for both assays with the lecithin-sphingomyelin ratio and phosphatidylglycerol. The correlations indicate that the recommended reference range for FLM will have more false predictions of immaturity than the NBD-phosphatidylcholine assay.
Medical subject headings
- Amniotic Fluid
- Fluorescence Polarization
- Lung
- Prenatal Diagnosis
- Pulmonary Surfactants