Local haemodynamics and shear stress in cuffed and straight PTFE-venous anastomoses: an in-vitro comparison using particle image velocimetry.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 14511997.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To use particle image velocimetry (PIV) to study the haemodynamics and shear stress associated with cuffed and straight PTFE-venous anastomoses. Silastic models of a straight and cuffed (Venaflo) PTFE-venous anastomoses were attached to a pulsatile flow 'Berlin Heart' circuit filled with glycerine/water and hollow glass tracer spheres. Instantaneous velocity fields were obtained PIV and shear rates and patterns calculated from frame-by-frame analysis. A high velocity jet struck the anastomotic 'floor' and was deflected toward the venous outflow. Shear stresses near the floor were significantly higher, in the straight anastomosis. Sites of high shear stress correlated well with the known sites of intimal hyperplasia. A cuffed anastomosis type may be favourable in terms of local haemodynamics so enhancing the long-term patency of PTFE-venous grafts.
Medical subject headings
- Anastomosis, Surgical
- Arteriovenous Shunt, Surgical
- Blood Vessel Prosthesis
- Hemodynamics
- Polytetrafluoroethylene
- Renal Dialysis