Fractures at the tip of long-stem prostheses used for revision hip arthroplasty.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 14513447.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We report a retrospective analysis of the incidence of intraoperative femur fractures at the tip of the prosthesis in 45 consecutive, uncemented, long (205-mm), straight hydroxyapatite, fully coated stems used for revision arthroplasty of the hip. Five patients were lost to follow-up. Intraoperative fractures of the distal femur occurred in 7 cases (17.5%). In the remaining 33 cases in which no fracture occurred, the tip of the implanted stem was found to be potentially at risk for causing a fracture in 18 (55%) cases. We concluded that uncemented, 205-mm, straight femoral stems should be used with caution in the setting of revision hip surgery. A bowed stem of that length may be a safer alternative. Otherwise, a shorter, straight stem may reduce the risk of complications at the stem tip.